Results with HIISI
Mixing gas: Oxygen
Microwave power: 2000 W + 300 W from 18 GHz + 14 GHz klystrons
Acceleration voltage:
(1)Tuning 28.6.2018
Same as (2) with following deviations: BD: -60 V, decreased argon and oxygen feed, decreased current in mid coil, increased injection B-field.
Acceleration voltage 18 kV/3.6 mA (HIISI extraction system). Plasma electrode aperture 8 mm in diameter.
(2) Tuning 27.6.2018 (test with recently assembled 36 segment hexapole). Background pressure in injection still quite high (1.5E-7 mbar).
Microwave power: 2000 W + 500 W from 18 GHz + 14 GHz klystrons. Fast tuning on Ar12+ at 19 kV. Intensity gradually increased when Ar feed was increased. At the total current of around 6 mA a strong spark appeared which destroyed at least puller HV power supply. I didn't have time to measure Ar13+ but spectrum indicates that it is about 330 µA or slightly more.
(3) Tuning 31.5.2018 about 10 % lower B-fields, 1 kW + 0.4 kW µw power
Acceleration voltage 16 kV/4 mA
Material consumption: N/A
Note: Tuning for Ar8+, Ar12+ or Ar13+ ion beam.
8+ | -600 (3) |
---|---|
9+ | -560 (3) |
10+ | - |
11+ | 400680(2) |
12+ | 462560 (2) |
13+ | 312 330 (2) |
14+ | 182195 (1) |
15+ | - |
16 | 31130 |
JYFL 14 GHz ECRIS
Mixing gas: Oxygen
...