Comment: This is the dedicated page to each laboratory. Add all values, pictures, information, links and comments you see relevant for the production of the ion beams. As an example, see JYFL Fe and Ti results. Modify the table in accordance with you CSD. Remove this comment after completion of the page).
Mixing gas:
Microwave power: W
Acceleration voltage: kV (extraction system)
Material consumption:
...
ECR4-14 GHZ :
The ion source is optimized on 40Ar9+ (black spectrum). This is a reference spectrum with 160 µA of Ar9+. After a plasma chamber replacement, the ion source is validated if the Ar9+ beam intensity is greater or equal to 120 µA. A record intensity of 245 µA has been reported on the graph below (red spectrum) although no clear conclusion could be made : much better distribution due to a magnesium wall coating effect, or due to the higher RF power injected into the plasma ?
Phoenix V2 - 18 GHz at SPIRAL 2 :
2016-04-18 15:02. Best Ar spectrum optimized on 40Ar14+ (46µA) obtained at SPIRAL2 with the Phoenix-V2 ion source.